This tutorial is part of the ID06 memo
Starting a compression
All of this is done in the Voggenreiter controller
- go into
Profile
and set a ramp, for instance
- 10 mn to reach 10 bars,
- 60 mn for 30 bars,
- and finally 2000 mn at 30 bars as final step
- go into
Settings
- check
Data storage
and that you have a .dat
log file
- make sure that the ckeckbutton
save
is selected
- go into
Controls
,
Active distance control
should be On
,
Hold follow
should be Hold position
,
Hold position ramp time
should be in agreement with compression times you chose in pofile
, 4200 sec for 70 mn, for instance,
End value
is the position that differentials rams should go to, to be determined with Jeremy.
- Go into
Profile
- Menu
PLC controller
Download profile to controller
- Menu
PLC controller
Run continue profile
- Go into
Controls
, activate Ramp On/Off
The press should be going up in pressure...
Start a deformation
If you are at pressure and want to start a deformation....
- Open the application:
press monitor
- Open the
Controls
tab
- Look at differential rams in
Control
tab
- choose
end value
and time
- example: remove 0.350 mm in 2h for 1 to 3.10-5 s-1
- Open the
Profile
tab
Suspend profile
in PLC Controller
menu
- remove about 2% of pressure on main ram over 2 hours, plus a long plateau afterward
Download profile
in PLC Controller
menu
Run profile
in PLC Controller
menu
- Back to differential rams in
Control
tab
Active distance control
should be On
,
Hold follow
should be Hold position
,
- Select
Ramp
to start deformation
You are good to go.