This tutorial is part of the ID06 memo
Starting a compression
All of this is done in the Voggenreiter controller
- go into
Profile and set a ramp, for instance
- 10 mn to reach 10 bars,
- 60 mn for 30 bars,
- and finally 2000 mn at 30 bars as final step
- go into
Settings
- check
Data storage and that you have a .dat log file
- make sure that the ckeckbutton
save is selected
- go into
Controls,
Active distance control should be On,
Hold follow should be Hold position,
Hold position ramp time should be in agreement with compression times you chose in pofile, 4200 sec for 70 mn, for instance,
End value is the position that differentials rams should go to, to be determined with Jeremy.
- Go into
Profile
- Menu
PLC controller
Download profile to controller
- Menu
PLC controller
Run continue profile
- Go into
Controls, activate Ramp On/Off
The press should be going up in pressure...
Start a deformation
If you are at pressure and want to start a deformation....
- Open the application:
press monitor
- Open the
Controls tab
- Look at differential rams in
Control tab
- choose
end value and time
- example: remove 0.350 mm in 2h for 1 to 3.10-5 s-1
- Open the
Profile tab
Suspend profile in PLC Controller menu
- remove about 2% of pressure on main ram over 2 hours, plus a long plateau afterward
Download profile in PLC Controller menu
Run profile in PLC Controller menu
- Back to differential rams in
Control tab
Active distance control should be On,
Hold follow should be Hold position,
- Select
Ramp to start deformation
You are good to go.